Efficient CMUT optimization using refracto-vibrometry

Sound at higher frequencies that is inaudible to the human ear is commonly referred to as “ultrasound.” This typically covers the acoustic spectrum from 20 kHz to about 1 GHz. Frequencies beyond this range are referred to as hypersound.

Nevertheless, ultrasound can be made visible to the human eye with the aid of optical measurement technology. Specifically, laser Doppler vibrometers (LDV) from Polytec can be used to make ultrasound visible.

To understand the mechanism behind this “audio-visual transduction,” it is necessary to understand how an LDV works. It is a so-called heterodyne laser interferometer, which is illustrated in Fig. 1 using a helium-neon laser source (wavelength 633 nm, red) as an example.

Fig. 1 | Setup of a laser Doppler vibrometer (LDV)

A laser beam is split into a measurement beam and a reference beam using a beam splitter. As the reference beam passes through an acousto-optic modulator (AOM, also known as a Bragg cell), its frequency is shifted upward by a precisely defined amount.

The frequency shift is only about 40 MHz and is therefore minimal given the fundamental frequency of the laser (e.g., 474 THz for red helium-neon-lasers). Nevertheless, for better visualization, the reference beam is shown in yellow in the figure.

The measurement beam, on the other hand, exits the LDV sensor head through a lens, and the signal reflected by the measurement sample (green) is captured again and directed, together with the reference beam, onto a photodiode via two additional beam splitters acting as mirrors. The resulting interference pattern of two waves with only minimally different frequencies generates a beat with a slowly oscillating amplitude. The photodiode can resolve this signal—which is perceptible as a fluctuation in brightness—relatively easily. In the static case, with the measurement sample at rest, the measured frequency then corresponds exactly to half of the frequency difference introduced by the Bragg cell (20 MHz).

In the standard mode of operation of the vibrometer (see Fig. 2), the measurement beam is reflected off a moving object and thus undergoes a frequency shift due to the optical Doppler effect—the faster the reflector moves toward the lens, the higher the frequency of the reflected beam. This reduces the frequency difference relative to the reference beam, and the brightness fluctuation at the photodiode slows down. Conversely, a reflector moving away from the lens causes a faster change in brightness. Finally, in the case of an oscillating reflector, its frequency manifests as a continuous alternation between a beat frequency that continuously increases and decreases. From this, one can infer the amplitude of the mechanical oscillation or the reflector’s vibration.

Fig. 2 | Laser Doppler vibrometer: Conventional mode (left) and refracto-vibrometry (right)

In the vibrometer’s operating mode used for ultrasound visualization, however, the reflector remains stationary. Nevertheless, a continuous sound wave travels through the air-filled space between the lens and the reflector. The speed of light is generally density-dependent. Because sound is a periodic density fluctuation, the time that the measurement beam requires to traverse the geometrically fixed distance between the lens and the reflector depends on the sound pressure within this space. Since both the measurement beam and the reference beam share the same time base (i.e., they are coherent with each other), the change in transit time causes a periodic phase shift, which mathematically has the same effect on the beat signal as in the conventional operating mode. This is why it is also referred to as a “virtual Doppler effect,” since the LDV detects an apparent vibration even though the reflector remains stationary.

Due to the relationship between density and refractive index, as well as the apparent vibration of the reflector caused by the sound propagating over it, this mode of operation of the interferometer is referred to as “refracto-vibrometry.” If a scanning LDV is used, it is ultimately possible to visualize entire sound fields.

This measurement technique was used in the laboratories at OTH Regensburg to measure the radiation patterns of capacitive ultrasonic transducers (CMUTs). A Polytec scanning vibrometer was used for this purpose.

CMUTs, as micromechanical components, have been the subject of research for many years, particularly for medical applications in ultrasound imaging. In addition, they are increasingly finding their way into other diagnostic and therapeutic as well as non-medical applications. OTH Regensburg focuses primarily on their use in the field of airborne ultrasound, for example as a sensor element for a flow meter.

Due to their inherently low susceptibility to structure-borne sound-induced crosstalk, CMUTs are particularly well-suited for the dense integration of multiple elements onto a single microchip. Such an array structure consisting of independently addressable transmitters thus enables so-called acoustic beamforming—that is, the electronically controlled manipulation of the sound’s radiation direction. Numerous applications thus benefit significantly from increased sensitivity of CMUTs or become feasible in the first place.

To verify the performance, it is necessary to record a high-resolution representation of the sound field of a CMUT at various beamforming angles. To this end, several transducers with frequencies ranging from 200 to 400 kHz were evaluated with respect to their radiation characteristics. Using refracto-vibrometry, complete sound fields could be scanned much more quickly than with a mechanically movable microphone probe. In particular, the intensity ratio of the main lobe to the side lobes is an important metric for beamforming quality.

For this purpose, an CMUT (a) with a center frequency of 266 kHz (corresponding to a wavelength in air under standard conditions of 1.29 mm) was fixed, as shown in Fig. 3, beneath an LDV measuring head (b) on an adjustment device (c) with a stationary reflector (d).

After the CMUT is excited, the measurement beam—focused on a single point—passes through the sound field twice. The sound-induced local pressure fluctuation alters the transit time of the laser beam only minimally. However, this is sufficient to detect an interferometric change in the measurement head. Once the measurement is set up, the Polytec Scanning Vibrometer records the two-dimensional sound field fully automatically.

Fig. 3 | Practical measurement setup for refracto-vibrometry

A representative selection of the results is shown in Fig. 4. Each sound field represents a kind of “snapshot” of the pressure distribution at a specific point in time.

Fig. 4 (a) | Point source (enlarged scale)
Fig. 4 (b) | Beam steering angle of 30°
Fig. 4 (c) | Beam steering angle of 45°
Fig. 4 (d) | Beam steering angle of 60°

In the first recording (a), a single CMUT element was excited. Because its size is smaller than the wavelength of the sound, the typical radiation pattern of a point source is observed. In contrast, the recordings (b)–(d) show the operation of the complete ultrasonic transducer array with various targeted beam steering angles. The respective main lobes in the desired directions, as well as the associated grating and side lobes, can be clearly identified.

In conclusion, refracto-vibrometry represents a highly precise and, above all, rapid method for characterizing ultrasonic transducers, particularly for small sound fields (< 5 x 5 cm²) and for wavelengths greater than the spot size of the laser at the focus. Looking ahead, this also makes the introduction of automated acoustic testing for mass production of ultrasonic transducers a viable prospect.

我们的作者

M.Sc. Sebastian Peller
Research Associate at the Sensor Application Center
sebastian.peller@oth-regensburg.de

相关文章

激光多普勒振动计在电动驱动系统的开发中树立了标杆

电动机正越来越多地被用作车辆的主要驱动技术。驱动系统的声学特性是影响电动汽车感知品质的关键因素。因此,在电动驱动系统的开发初期,就会采用数值方法来评估和优化其声学特性。

5 min

光学自由曲面制造工艺的优化

通过采用光学自由曲面,可以简化复杂的透镜系统,从而使其结构更紧凑、重量更轻。此外,光学自由曲面的应用还能实现完全崭新的功能。但与此同时,其制造工艺较为复杂,因此耗时且成本高昂。

4 min

关于自支撑和自对中型紧凑型振动筛

人工激励是各种噪声与振动技术的基础,涵盖从实验模态分析、传递路径分析、载荷识别、仿真模型关联到混合建模等多个方面。结构激励可通过多种方式实现,包括手动冲击锤、自动冲击锤、外支撑振动台以及自支撑振动台等。

10 min

基于激光振动计的声学性能评估(-40°C 至 85°C)

随着触摸屏界面在汽车应用中的普及,信息视觉负荷日益加重,这给行车安全带来了严重影响。触觉反馈技术被视为改善用户体验、减轻触摸屏交互认知负荷的一种途径。

5 min

用于模态试验的惯性振动台

如果想要测试一个本身不产生振动物体的振动特性,就必须对其进行外部激励。在模态试验中,电磁振动台和模态锤都是常见的宽带激励工具,多年来已被广泛应用。……

6 min

如何测量非洲象的低频声学通讯

最近发表在《欧洲物理杂志·专题》上的一项研究,介绍了一项将激光多普勒振动测量法(LDV)应用于深入理解非洲象发声及其行为意义的开创性研究。该研究的主要目的是确定发声互动的动态过程……

4 min

“会唱歌”的电容器如何影响消费电子产品的音频体验

无论是在移动设备上观看节目、与亲友交流,还是在商务场合,大家都喜欢享受优质的音质。反之,一旦音频出现问题,我们就会感到不舒服。……

3 min

剪切冲击波如何影响感知

触觉全息技术或空中触觉技术有望让虚拟现实栩栩如生,但一项新研究揭示了一个需要克服的令人惊讶的物理障碍。加州大学圣塔芭芭拉分校的一个研究团队发现了一种新现象……

6 min

分析扬声器的工作振形

噪声防护与控制在专业和私人环境中都起着重要作用。作为一种标准测试方法,空气传声测量有助于识别声传播路径,从而实现降噪。通常,通过使用全向声源来产生测试所需的声源……

4 min

基于声波晶体(SAW)的声流体技术对复杂流体的操控

基于声表面波(SAW)的声流体学将微流体技术与有源微声场相结合。这使得能够实现多种“芯片实验室”操作,涵盖从流体操控(包括混合、液滴和气溶胶生成)到细胞及(生物)颗粒的操控。

3 min

声流体学*的一种新型驱动机制

声流体学是指在微米和纳米尺度上对流体以及浸没其中的颗粒或细胞进行主动操控,是生命科学领域的一项关键技术。声波——更确切地说,是频率远超人耳可听范围的超声波——天生就适合用于这一目的……

5 min

激光振动测量技术实现触觉反馈的可视化

智能手机已经证明,在不牺牲使用舒适性的前提下,可以取代实体按键。支持手势控制且可自由配置的显示屏,让应用程序之间的切换变得更加轻松。Hap2U正在推动这场人机交互革命向前发展。其创新技术提供了高保真的触觉反馈……

7 min
加载更多

下載

請填寫表格,您將在幾分鐘內透過電子郵件收到下載連結。