Exciting comparison

Inertial shakers for modal testing

State-of-the-art excitation methods

If you would like to test the vibration behavior of an object that is not vibrating by its own, you have to excite it externally. For modal tests both an electromagnetic shaker and a modal hammer are common broadband excitation tools that has been widely used for many years. For a measurement using a Polytec Scanning Laser Vibrometer, which measures points sequentially, a hammer impact has to be repeated for every scan point. For this reason, automated modal hammers like the SAM, designed by NV-Tech, have become popular in recent years.

Both existing excitation methods have strengths and weaknesses. An electrodynamic shaker leads to faster measurements over many points and repeatable test results. But a shaker with force sensor can have a non-neglectable influence on the structure under test. And, in practice, setting up and aligning the shaker, maintaining alignment, is cumbersome, especially on test setups where the sample is supported by soft rubber strings or soft foam to mimic free-free conditions.

An (automated) hammer on the other side has ideally no impact on the inherent dynamic properties of the part, making it a good choice for relatively lightweight and lightly damped structures. Due to the decay time after each hit the measurements take considerably more time and, for lightweight samples suspended by soft rubber strings, might lead to considerable swaying of the entire sample.

Combining strengths

Recently, a third excitation method has appeared tackling the shortfalls of the above-mentioned methods: Inertial shakers with integrated force transducers from Qsources. These shakers have some interesting features like a self-aligning internal suspension, very low mass and an easy handling and alignment. Being attached to the structure the position is fixed and they avoid the waiting time with automated hammers. On the downside, they only work well above a certain cutoff frequency and load the structure in a similar way as an accelerometer with cable would. For this article, we have tested the Qlws inertial shaker and compared it to a classical shaker and an automated modal hammer.

Use of the Qlws light weight inertial shaker

The Qlws itself consists of 2 small parts (+ an amplifier):

  • a so-called “force cup” contains the voice coil. It is glued to the object under test.
  • the shaker body, containing the inertial mass and a force sensor. It is simply manually pushed over the force cup.

Between the two, a thin layer of grease is used for dynamic decoupling, leading to a dynamically active mass of < 2 g  in the range above 200 Hz.

Attaching the Qlws is easy: clean the sample with alcohol, glue the force cup with instant glue, put a thin layer of grease, slide the shaker body over the force cup and connect the cables to the amplifier box and to a data acquisition featuring IEPE power supply.

No stinger alignment is necessary, as the device is self-aligned in any inclination.

Figure 1: The Qlws components

Attaching the Qlws is easy: clean the sample with alcohol, glue the force cup with instant glue, put a thin layer of grease, slide the shaker body over the force cup and connect the cables to the amplifier box and to a data acquisition featuring IEPE power supply.

No stinger alignment is necessary, as the device is self-aligned in any inclination.

Figure 2: The Qlws inertial shaker mounted to the DUT

Comparison test

Comparison has been done on a 1,5 mm thick aluminum plate, suspended by elastic straps. This plate is excited by a classical modal shaker, an automated hammer “SAM 1” and the Qsources Qlws. To measure the vibration behavior, a Polytec non-contact PSV Laser Scanning Vibrometer is used which has no influence on the test object.

The following images show the electrodynamic shaker (LDS shaker), the automatic hammer (NV-Tech SAM), and the inertial shaker (Qsources Qlws):

Automatic hammer (NV-Tech SAM)
Electrodynamic shaker (LDS shaker)
Inertial shaker (Qsources Qlws)

The LDS shaker is quickly seen having too much influence on the light test object: resonance frequencies are shifted, and the force spectrum has some dramatic drops, leading to bad signal to noise ratio at the transfer function between response and excitation. The conventional modal shaker is therefore not suitable for this setup and should be used for heavier samples only where it has its strengths.

The NV-Tech SAM automated hammer is the reference in terms of mass loading of the test-object since nothing is permanently attached to the structure. It is therefore the excitation source of choice for objects with very small damping. Its frequency range is also extending to low values. For the very light suspended test sample, it has nevertheless a drawback: It leads to swaying of the suspended sample. This, together with the use of a scanning vibrometer leads to unnecessary noise in the measurements. This effect can to some extent be mitigated by the use of the latest Polytec QTec® technology in the vibrometer, but the signal to noise ratio will still be somewhat reduced by the macroscopic swaying. The swaying also increases the test time, as the user has to wait before the next automated hit can be applied. If swaying can be avoided (e.g., if sample support on soft foam is acceptable), the hammer is still a very good choice for objects with little damping and for a wide range of frequencies.

The Qsources Qlws has only very limited influence on the natural behavior of the sample due to its small dynamic mass. If placed correctly, the influence of coupling the shaker to the test-object eigenfrequencies is very small, comparable to the effect of suspending the test-object, for frequencies below 2 kHz, which covers the first 15 - 20 resonances. No waiting time between excitation sequences is required, therefore speeding up the test considerably. The only drawback that appeared during the test is the low frequency limitation below 250 Hz. It does not mean that the shaker is not working for lower frequencies , but at lower frequencies it has slightly more influence on the natural behavior of the sample due to the higher effective dynamic mass.

 

Figure 3 shows typical results of SAM hammer (left) and Qlws (right) for this setup.

 

The main results are:
 

  • Resonance frequencies and corresponding deflection shapes are virtually identical between 200 and 2,000 Hz.
  • Below 200 Hz, the inertial shaker only has some minor influence, also above 2 kHz.
  • The increase in noise due to the swaying, caused by the hammer impact, can be clearly seen in the depicted deflection shape.
  • The scan time for the measurement with the automated hammer was about 2.5 h as compared to 0.5 h using the Qlws.
  • And, even a small conventional electrodynamic shaker with stinger and force sensor is not suitable for this type of samples.
Figure 3: Deflection shapes and FRF (left: modal hammer, right: Qlws)

Resume

To resume, the Qsources Qlws shaker is a very interesting alternative to an automated modal hammer, as it produces no swaying of the test object which lowers the signal to noise ratio for single channel Vibrometers without the Polytec QTec® technology.

Since this additional noise can be compensated using multi-channel QTec® Scanning Vibrometers both technologies, the inertial shaker and the automated hammer produce test data on the same level. Limitations can be the lower frequency limit of the Qlws and the added mass for extremely light weight or extremely low damped samples.

The setup of the Qlws is very convenient, comparably simple as an automated hammer, has excellent repeatability, and is a lot faster than the hammer and easier than a classical shaker with stinger. As with the Qlws there is no waiting time between hits, the measurement time can be considerably shorter for suspended samples.

Images courtesy: Images courtesy of the authors unless otherwise specified.

Our Authors

Dr. Jochen Schell
Manager Application Services Business Unit Vibrometry, Polytec GmbH, Germany
info@polytec.de

Related Articles

Laser-Doppler-Vibrometer set standards in the development of electric drives

Electric machines are increasingly being used as the primary drive technology in vehicles. The acoustic behavior of the drive is a key factor in the perceived quality of electric vehicles. Numerical method…

5 min

Optimized manufacturing processes for optical free-form surfaces

The use of optical free-form surfaces allows complex lens systems to be simplified and thus made more compact and weight-optimized. In addition, the use of optical freeform surfaces enables the realization…

4 min

On self-supporting and self-aligning compact shakers

Artificial excitation is the basis for various noise and vibration techniques, from experimental modal analysis, over transfer path analysis, load identification, simulation model correlation, hybrid model…

10 min

Focus on impact detection of micrometeorites

The commercialization of space exploration increases the need for innovative technologies for space and planetary exploration and satellite missions. To ensure the safety and reliability of space structure…

6 min

Laser vibrometer based acoustic performance evaluation from -40° to 85°C

The increase of touchscreen interfaces in automotive applications leads to a visual overload of information that has serious safety implications. Haptic feedback technology is seen as an opportunity to imp…

5 min

How to measure low-frequency acoustic communication of African elephants

A recent study published in the European Physical Journal Special Topics introduces a pioneering application of laser Doppler vibrometry (LDV) to better comprehend African elephant vocalizations and their …

4 min

How singing capacitors affect the audio experience in consumer electronics

Everybody enjoys good sound quality on a mobile device when watching a show or when communicating with friends and family or in a business environment. On the other hand, we cringe when something is off wi…

3 min

When to choose a laser Doppler vibrometer for vibration measurement

In the realm of vibration testing, a significant obstacle arises in the form of mass loading. The behavior of a dynamic vibration sample, or device under test (DUT), is profoundly influenced by the additio…

5 min

Single electrode transduction of contour modes in piezoelectric disk resonators

Piezoelectric disk resonators are designed to transduce specific contour mode bulk acoustic wave (BAW) resonant shapes. Utilizing a unique property of lead magnesium niobite-lead titanate (PMN-PT) thin fil…

4 min

How shear shock waves influence perception

Haptic holography or mid-air haptics promises to bring virtual reality to life, but a new study reveals a surprising physical obstacle that will need to be overcome. A research team at UC Santa Barbara has…

6 min

Celebrating the iconic career of Professor Rong Z. Gan

We had a chance to virtually sit down for an interview with Prof. Rong Zhu Gan. In her field, she is one of those personalities that needs no introduction. For the uninitiated, she is the George Lynn Cross…

3 min

Boundary conditions, measurement inaccuracy, and repeatability

MAN Energy Solutions offers a comprehensive engine portfolio, propeller, and stern equipment and turbochargers for the maritime sector. Reliability is an essential aspect of our sector and especially vital…

5 min

A modal-based monitoring system

More and more, the increasing stress on infrastructures provides the challenge of investigating and evaluating the existing structures with regard to safety and the resulting remaining useful life. Reliabl…

5 min

Non-contact alternative to dynamic and viscoelastic mechanical testing in cartilage

Osteoarthritis, a degenerative disease in the joint, is a condition that is afflicting approximately 14% of adults. Understanding the dynamics of the joint cartilage is crucial for research and for finding…

6 min

Frequency response function measurement using QTec® multi-path interferometry

The measurement of frequency response function (FRF) plays an important role for the dynamic characterization of structures. In most cases, the goal is the measurement-based analysis of a system or the det…

7 min

Analyzing operational deflection shapes on loudspeakers

Noise protection and control play a major role in both professional and private environments. Airborne sound measurements help identify acoustic paths as a means of noise reduction as standard testing meth…

4 min

Manipulation of complex fluids with SAW-based acoustofluidics

Surface acoustic wave (SAW)-based acoustofluidics combines microfluidics with active microacoustic fields. This allows to realize numerous Lab-on-a-Chip operations, reaching from fluid manipulation includi…

3 min

A new driving mechanism for acoustofluidics*

Acoustofluidics is the active manipulation of fluids as well as of immersed particles or cells at the micro- and nanoscale and is a key technology for life sciences. Acoustic waves, more precisely, ultraso…

5 min

Laser vibrometry visualizing haptic feedback

The smartphone has shown that buttons can be replaced without sacrificing comfort. The freely configurable display with gesture control has made it even easier to switch between applications. Hap2U drives …

7 min

Noise and modal analysis for quiet and reliable operation

The growth of the drone market is multi-faceted: growing concerns about national and border security drive the need for advanced aerial surveillance, package delivery services experience an increased deman…

6 min

Measurements using broadband piezoelectric excitation

The following article compares various forms of force excitation in order to determine the transfer admittances of mechanical systems using a gear wheel as an example. The objective is to analyze and asses…

4 min

Determining the dynamic characteristics of an aircraft’s large-scale, adaptive compliant wing slat section prototype

The SADE project studied “smart” morphing mechanical elements of next generation wings, which was aimed at improving aerodynamic performance, fuel efficiency, as well as reducing noise and emissions in all…

3 min

Acoustic radiation of vibrating surfaces in electric drive systems

The growing electric and hybrid vehicles market faces new demands on product acoustic quality. The acoustical emission behavior of electric drivetrains is clearly different from that of conventional drives…

4 min

Vibration analysis helps to protect art

To organize traveling exhibitions, museums like the Georgia O’Keeffe Museum in Santa Fe, New Mexico supplement their own collection with temporary loans from other museums and private collections. In 2017,…

4 min

Experimental modal test with laser vibrometers

The validation of simulation models is an important application area of laser vibrometers. To do this, the geometric data from a finite element mesh is loaded into the software of the measurement system an…

4 min

3D scanning laser Doppler vibrometry for fuel-efficient turbines

Fuel consumption of aircrafts is a hot topic in the current discussion about global climate change. An important factor is the design of the aircraft engine itself. In order to reduce weight and build more…

6 min

Xtra 3D scanning vibrometry on rolling tires

Dynamic 3D operational deflection shapes captured with an Xtra Scanning Vibrometer help reveal the source of rolling tire noise. For a wide range of driving conditions, rolling tires stimulate a significan…

5 min
Load More